R&S FSL SENSe Subsystem (Noise Figure, K30)
1300.2519.12 6.553 E-11
Example
CORR:LOSS:OUTP:MODE TABL
Uses the loss output table.
CORR:LOSS:OUTP:TABL 1MHz,10,2MHz,12
Specifies a new output loss table and overwrites the current output loss table with the two entry
pairs specified.
C
haracteristics
*RST value: –
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:TEMPerature
This command specifies the room temperature of the operating environment. This value is taken
into account when calculating noise results.
Parameter
278.15 to 318.15 K; up to 2 decimal places can be specified.
Example
CORR:TEMP 291.50
Specifies the room temperature to 291.50 Kelvin (18.5 C).
Characteristics
*RST value: 293 K
SCPI: device–specific
Mode
NF