SENSe Subsystem (Noise Figure, K30) R&S FSL
1300.2519.12 6.552 E-11
[SENSe:]CORRection:LOSS:OUTPut:MODE
This command specifies whether a constant output loss value applies for all measured
frequencies or an output loss list (loss output values specified at specific input frequencies). The
constant loss output value is specified by the [SENSe:]CORRection:LOSS:OUTPut:SPOT
command.
Parameter
SPOT The constant loss input value for all measurement frequencies is used.
TABLe The loss input table is used.
Example
CORR:LOSS:OUTP:MODE SPOT
Uses the configured constant loss output value for all input frequencies to be measured.
Characteristics
*RST value: SPOT
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:LOSS:OUTPut:SPOT
This command specifies the loss output constant for all input frequencies to be measured.
Parameter
–999.99 to 999.99 dB, accurate to two decimal places
Example
CORR:LOSS:OUTP:MODE SPOT
Uses the configured constant loss output value for all input frequencies to be measured.
CORR:LOSS:OUTP:SPOT 10
Sets the internal output loss constant value to 10 dB for all input frequencies to be measured.
Characteristics
*RST value: 0 dB
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:LOSS:OUTPut:TABLe
This command specifies a new output loss table to determine the correct output loss to be used
for the input frequencies. This new list completely overwrites all current output loss list entries
regardless of how many entries are present and how many entries are being supplied for the
new list.
Parameter
<numeric_value>,<numeric_value>,... (maximum of 100 argument pairs)
<numeric_value>,<numeric_value> = a frequency loss pair of arguments
frequency: 0 Hz to 999.99 GHz (max. two decimal places)
loss figure: –999.99 to 999.99 (value in dB, accurate to two decimal places)