EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1359

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands of the CDMA2000 BTS Analyzer Option (K82) R&S FSL
1300.2519.12 6.600 E-11
Example
CALC:LIM:ESP:REST
Resets the limit lines for the Spectrum Emission Mask to the default setting.
Characteristics
RST value: -
SCPI: device-specific
Mode
CDMA
CALCulate<1|2>:LIMit<1...8>:ESPectrum:VALue
This command activates the manual limit line selection and specifies the expected power as a
value. Depending on the entered value, one of the predefined limit lines is selected.
The numeric suffix at CALCulate<1|2> specifies the measurement window. The numeric suffix
at LIMit<1...8> is irrelevant for this command.
Parameter
33
P
33
28
28
P < 33
0 P < 28
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CONF:CDP:BCL 1
Select band class 1, 1900 MHz
CONF:CDP:MEAS ESP
Selects Spectrum Emission Mask measurement.
CALC:LIM:ESP:VAL 33
Activates manual selection of the limit line and selects the limit line for P
33.
INIT;*WAI
Starts measurement with synchronization.
CALC:LIM:FAIL?
Queries result of the limit check.
Characteristics
RST value: 0
SCPI: device-specific
Mode
CDMA

Table of Contents