R&S FSL CALCulate Subsystem (CDMA, K82)
1300.2519.12 6.601 E-11
CALCulate:MARKer<1…4>:FUNCtion Subsystem (CDMA, K82)
The CALCulate:MARKer:FUNCtion subsystem checks the marker functions in the instrument.
Commands of the CALCulate:MARKer<1…4>:FUNCtion Subsystem
– CALCulate<1|2>:MARKer<1...4>:FUNCtion:CDPower[:BTS]:RESult?
– CALCulate<1|2>:MARKer<1...4>:FUNCtion:PICH
– CALCulate<1|2>:MARKer<1...4>:FUNCtion:TDPIch
CALCulate<1|2>:MARKer<1...4>:FUNCtion:CDPower[:BTS]:RESult?
This command queries the results of the code domain measurement for the selected channel.
The channel is selected via the [SENSe<1|2>:]CDPower:CODE command.
The numeric suffix at CALCulate<1|2> specifies the measurement window. The numeric suffix
at MARKer<1...4> is irrelevant for this command.
Parameter
SLOT | PTOTal | PPICh | RHO | MACCuracy | PCDerror | ACTive | FERRor | FERPpm |
CERRor | TFRame I IQOFfset | IQIMbalance | SRATe | CHANnel | SFACtor | TOFFset |
POFFset | CDPabsolute | CDPRelative | EVMRms | EVMPeak | DMType
For a description of the parameter refer to "Result Summary" on page 6.629.
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
INIT;*WAI
Starts measurement with synchronization.
CALC:MARK:FUNC:CDP:RES? PTOT
Reads out total power.
CDP:SLOT 2
Selects power control group 2.
CDP:CODE 11
Selects code number 11.
CALC:MARK:FUNC:CDP:RES? EVMR
Reads out EVM rms of the code with number 11 in PCG 2.
Characteristics
RST value: -
SCPI: device-specific
Mode
CDMA
CALCulate<1|2>:MARKer<1...4>:FUNCtion:PICH
This command sets the marker to channel 0.64.
The numeric suffix at CALCulate<1|2> specifies the measurement window. The numeric suffix
at MARKer<1...4> specifies the marker to be set to the pilot channel.
This command is an event and therefore has no *RST value and no query.