Remote Commands of the CDMA2000 BTS Analyzer Option (K82) R&S FSL
1300.2519.12 6.616 E-11
Characteristics
RST value: OFF
SCPI: device-specific
Mode
CDMA
[SENSe<1|2>:]CDPower:CODE
This command selects the code number.
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
0 to 127; the maximum value depends on the base spreading factor;
For further details refer to chapter 4, "Channel (Code) Number".
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:SFAC 128
Selects base spreading factor 128.
CDP:CODE 11
Selects code number 11.
INIT;*WAI
Starts measurement with synchronization.
Characteristics
RST value: 0
SCPI: device-specific
Mode
CDMA
[SENSe<1|2>:]CDPower:ICTReshold
This command defines the minimum power which a single channel must have compared to the
total signal in order to be regarded as an active channel. Channels below the specified
threshold are regarded as "inactive".
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
–100 dB to 0 dB
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:ICTR –10DB
Sets threshold value to –10 dB.
INIT;*WAI
Starts measurement with synchronization.