Remote Commands of the 1xEV-DO BTS Analyzer Option (K84) R&S FSL
1300.2519.12 6.650 E-11
Example
CONF:CDP:MEAS POW
Selects Signal Channel Power measurement .
Characteristics
RST value: CDPower
SCPI: device-specific
Mode
EVDO
CONFigure:CDPower[:BTS]:PVTime:BURSt
This command activates an automatic burst alignment to the center of the diagram.
Parameter
ON | OFF
Example
CONF:CDP:PVT:BURS ON
Activates the automatic alignment
Characteristics
RST value: OFF
SCPI: device-specific
Mode
EVDO
CONFigure:CDPower[:BTS]:PVTime:FREStart
If switched on, this command evaluates the limit line over all results at the end of a single
sweep. The sweep is restarted if this result is FAILED.
Parameter
ON | OFF
Example
CONF:CDP:PVT:FRES ON
Restarts a single sweep if the result evaluation is failed.
Characteristics
RST value: OFF
SCPI: device-specific
Mode
EVDO