R&S FSL CALCulate:MARKer Subsystem (WLAN, K91 / K91n)
1300.2519.12 6.695 E-11
Characteristics
*RST value: –
SCPI: device–specific
Mode
WLAN
CALCulate<1|2>:MARKer<1>:Y
This command positions the selected marker to the indicated quadrature (Constellation vs
Symbol), magnitude of I or Q (Constellation vs Carrier), EVM (EVM vs Carrier) or abs (Spectrum
Flatness) in the selected measurement window.
This command is query only for the following result displays:
Auto Level
Constellation vs Symbol
Constellation vs Carrier
EVM vs Symbol
PVT Full
PVT Rising / Falling
Magnitude Capture Buffer
Spectrum Mask
Spectrum ACP/ACPR
Spectrum FFT
CCDF
Parameter
<numeric_value> in percent or dB
Example
CALC2:MARK:Y –2
Positions marker 1 in screen B to –2.
CALC:MARK:Y?
Outputs the measured value of marker 1.