CALCulate:MARKer Subsystem (WLAN, K91 / K91n) R&S FSL
1300.2519.12 6.696 E-11
CALCulate:MARKer:FUNCtion Subsystem (WLAN, K91 / K91n)
The measurement window is selected by CALCulate 1 (screen A) or 2 (screen B).
Commands of the CALCulate:MARKer:FUNCtion Subsystem
CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult[:CURRent]?
CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult:MAXHold?
CALCulate<1|2>:MARKer<1>:FUNCtion:ZOOM
CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult[:CURRent]?
This command queries the current result values of the adjacent channel power measurement.
An ACPR (Adjacent channel power relative) measurement must have previously been run, for
there to be summary data available.
Results are output separated by commas in the following order:
Power of main channel
Power of lower adjacent channel
Power of upper adjacent channel
Power of lower alternate adjacent channel 1
Power of upper alternate adjacent channel 1
Power of lower alternate adjacent channel 2
Power of upper alternate adjacent channel 2
Adjacent channel power values are output in dB.
This command is a query only and thus has no *RST value.
Example
CALC:MARK:FUNC:POW:RES?
Characteristics
*RST value: –
SCPI: device–specific
Mode
WLAN
CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult:MAXHold?
This command queries the maximum result values of the adjacent channel power
measurement. An ACPR (Adjacent channel power relative) measurement must have previously
been run with more than one sweep, for there to be maximum summary data available.
For details on the output refer to
"CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult[:CURRent]?" on page 6.696.
This command is a query only and thus has no *RST value.
Example
CALC:MARK:FUNC:POW:RES:MAXH?