CONFigure Subsystem (WLAN, K91 / K91n) R&S FSL
1300.2519.12 6.706 E-11
CONFigure:POWer:AUTO
This remote control command is used to switch on or off automatic power level detection. When
switched on, power level detection is performed at the start of each measurement sweep.
Parameter
ONCE | ON | OFF
If this command is issued with the ONCE parameter then the auto level routine is performed
immediately one time.
Example
CONF:POW:AUTO ON
Configures the automatic detection of the input power level.
Characteristics
*RST value: –
SCPI: device–specific
Mode
WLAN
CONFigure:POWer:AUTO:SWEep:TIME
This remote control command is used to specify the sweep time for the automatic power level
detection.
Parameter
numeric value in seconds
Example
CONF:POW:AUTO:SWE:TIME 200MS
Configures a 200 ms sweep time for the auto–level detection.
Characteristics
*RST value: 100ms
SCPI: device–specific
Mode
WLAN
CONFigure:POWer:EXPected:RF
This remote control command is used to specify the mean power level of the source signal as
supplied to the Analyzer RF input. This value will be overwritten if Auto Level is turned on.
Parameter
<numeric_value> in dBm
Example
CONF:POW:EXP:RF 9
Assumes an input signal strength of 9 dBm.
Characteristics
*RST value: –
SCPI: device–specific
Mode
WLAN