EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1555

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONFigure Subsystem (WiMAX, K92/K93) R&S FSL
1300.2519.12 6.796 E-11
CONFigure:POWer:AUTO
This command switches on or off the automatic power level detection.
Parameter
ON | OFF | ONCE
Example
CONF:POW:AUTO ON
At the start of every measurement sweep the input power level is detected automatically.
Characteristics
RST value:
SCPI: device–specific
Mode
OFDM, OFDMA/WiBro
CONFigure:POWer:AUTO:SWEep:TIME
This command specifies the sweep time for the automatic power level detection.
Parameter
1ms to 1s
Example
CONF:POW:AUTO ON
At the start of every measurement sweep the input power level is detected automatically.
CONF:POW:AUTO:SWE:TIME 200MS
The sweep time is set to 200 ms power level.
Characteristics
RST value: 100 ms
SCPI: device–specific
Mode
OFDM, OFDMA/WiBro
CONFigure:POWer:EXPected:RF
This command specifies the input power level of the source signal that will be supplied at the
analyzer RF input.
Parameter
–999.99 dBm to 999.99 dBm
Example
CONF:POW:EXP:RF –20
Assumes an input signal strength of –20 dBm.
Characteristics
RST value: –30 dBm
SCPI: device–specific