R&S FSL DISPlay Subsystem (WiMAX, K92/K93)
1300.2519.12 6.815 E-11
D
ISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:RLEVel:OFFSet
This command specifies the external attenuation/gain applied to measurements. The value
corresponds to the reference level offset in spectrum analyzer mode (see
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RLEVel:OFFSet command).
T
he numeric suffixes of WINDow<1|2> and TRACe<1 to 3> are irrelevant.
Parameter
–200 to 200 dB
Example
DISP:TRAC:Y:RLEV:OFFS 10
External attenuation (level offset) of the analyzer is 10 dB
DISP:TRAC:Y:RLEV:OFFS –10
External attenuation of the analyzer is –10 dB. i.e. a gain of 10 dB
Characteristics
RST value: 0 dB
SCPI: conform