R&S FSL Setting Traces – TRACE Key
1300.2519.12 4.45 E-11
Average
Selects the Average mode. For details see "Trace mode overview" on page 4.40.
This softkey is available from firmware version 1.80.
Remote: DISP:TRAC:MODE AVER
View
Selects the View mode. For details see "Trace mode overview" on page 4.40.
This softkey is available from firmware version 1.80.
Remote: DISP:TRAC:MODE VIEW
Blank
Selects the Blank mode. For details see "Trace mode overview" on page 4.40.
This softkey is available from firmware version 1.80.
Remote: DISP:TRAC OFF
Detector Auto Select
Selects the optimum detector for the selected trace and filter mode. This is the default setting.
For details see also "Detector overview" on page 4.42.
Trace mode Detector (band–pass filter) Detector (FFT filter)
Clear/Write Auto Peak Max Peak
Average Sample Sample
Max Hold Max Peak Max Peak
Min Hold Min Peak Max Peak
Remote: DET:AUTO ON
Detector Manual Select
Opens a submenu to select the detector. For details see "Detector overview" on page 4.42.
Detector Auto Peak
Selects the Auto Peak detector. For details see "Detector overview" on page 4.42.
This softkey is available from firmware version 1.80.
Remote: DET APE