Setting Traces – TRACE Key R&S FSL
1300.2519.12 4.48 E-11
Decim Sep
Selects the decimal separator with floating–point numerals for the ASCII file export to support
evaluation programs (e.g. MS Excel) in different languages. The values '.' (decimal point) and ','
(comma) can be set. For details see also ASCII File Export softkey.
Remote: FORM:DEXP:DSEP POIN
Trace Math
Opens the Trace Mathematics dialog box to define which trace is subtracted from trace 1. The
result is displayed in trace 1 and refers to the zero point defined with the Trace Math Position
softkey. The following substractions can be performed:
T1–>T1–T2
Substracts trace 2 from trace 1.
T1–>T1–T3
Substracts trace 3 from trace 1
T1–>T1–T4
Substracts trace 4 from trace 1
T1–>T1–T5
Substracts trace 5 from trace 1
T1–>T1–T6
Substracts trace 6 from trace 1
If the Trace Math Off option is activated, the function is switched off (default setting).
This softkey is available from firmware version 1.30.
Remote: CALC1:MATH (TRACE1 – TRACE2)
Remote: CALC:MATH:STAT ON
Trace Math Position
Opens an edit dialog box to define the zero point in % of the diagram height. The range of
values extends from –100% to +200%.
This softkey is available from firmware version 1.30.
Remote: CALC:MATH:POS 50PCT
Average Mode
Opens a submenu to select the averaging method for the average trace mode. The submenu
contains the following softkeys.
This softkey is available from firmware version 1.90.
Command
LOG
LIN
POWER