Power Measurements – MEAS Key R&S FSL
1300.2519.12 4.108 E-11
S
top after Sweep (Sweep List dialog box, Spurious Emissions)
Configures the sweep behavior.
On
The R&S FSL stops after one range is swept and continues only if you confirm (a message
box is displayed).
Off
The R&S FSL sweeps all ranges in one go.
Remote: LIST:RANG1:BRE ON
Transd. Factor (Sweep List dialog box)
Sets a transducer for the specified range. You can only choose a transducer that fulfills the
following conditions:
– The transducer overlaps or equals the span of the range.
– The x–axis is linear.
– The unit is dB.
Remote: ESP:RANG1:TRAN 'test' (Spectrum Emission Mask)
Remote: LIST:RANG1:TRAN 'test' (Spurious Emissions)
Limit Check 1 to 4 (Sweep List dialog box)
Sets the type of limit check for all ranges. Possible states are:
Absolute
Checks only the absolute limits defined.
Relative
Checks only the relative limits. Relative limits are defined as relative
to the measured power in the reference range.
Abs and
Rel
Combines the absolute and relative limit. The limit check fails when
both limits are violated.
Abs or Rel
Combines the absolute and relative limit. The limit check fails when
one of the limits is violated.
The limit state affects the availability of all limit settings (Abs Limit Start, Abs Limit Stop, Rel
Limit Start, Rel Limit Stop).
Dependent on the number of active power classes (see Edit Power Classes dialog box), the
number of limits that can be set varies. Up to four limits are possible. The sweep list is extended
accordingly.
Remote: ESP:RANG3:LIM:STAT AND (Spectrum Emission Mask)
Remote: LIST:RANG3:LIM:STAT ON (Spurious Emissions)
Remote: CALC:LIM3:FAIL?