R&S FSL Using Limit Lines and Display Lines – LINES Key
1300.2519.12 4.125 E-11
Edit Margin
Sets the focus on the Margin field to enter or change a margin for the limit line. The default
setting is 0 dB (i.e. no margin).
This softkey is available from firmware version 1.10.
Value
Opens an edit dialog box to change an existing x or y value, depending on the selected column.
The softkey is only available if an existing value is selected.
The desired data points are entered in ascending order (two repeated frequencies/time values
are permitted).
This softkey is available from firmware version 1.10.
Remote: CALC:LIM3:CONT:DATA 1MHz,3MHz,30MHz
Remote: CALC:LIM3:UPP:DATA –10,0,0
Remote: CALC:LIM3:LOW:DATA –30,–40,–40
Insert Value Above
Creates an empty line above the selected data point to enter a new data point. This softkey
corresponds to the Insert button in the dialog box.
It is also possible to add a data point at the end of the list, if the focus is set below the last entry
line of the list.
The data points are entered in ascending order (two repeated frequencies/time values are
permitted). If the entered values are not in accordance with the ascending order rule, an error
message is displayed and the values are discarded.
This softkey is available from firmware version 1.10.
Delete Value
Deletes the selected data point (x and y value). All succeeding data points are shifted up
accordingly. This softkey corresponds to the Delete button in the dialog box.
The softkey is only available if an existing value is selected.
This softkey is available from firmware version 1.10.
Save Limit Line
Saves the currently edited limit line under the name defined in the Name field.
This softkey is available from firmware version 1.10.