Using Limit Lines and Display Lines – LINES Key R&S FSL
1300.2519.12 4.126 E-11
Edit
Opens a submenu to edit limit lines. For details see also "Limit Lines" on page 4.123 and "To
edit an existing limit line" on page 4.121.
This softkey is available from firmware version 1.10.
Remote: For details refer to chapter "Remote Control – Commands", section "Definition of the limit
line".
Copy to
Copies the data of the selected limit line and displays it in the Edit Limit Line dialog box. If the
limit line is edited and saved under a new name, a new limit line can be easily generated by
parallel translation or editing of an existing limit line.
For details see also "Limit Lines" on page 4.123 and "To create a new limit line based upon an
existing limit line" on page 4.121.
This softkey is available from firmware version 1.10.
Remote: CALC:LIM3:COPY 2
Delete
Deletes the selected limit line.
This softkey is available from firmware version 1.10.
Remote: CALC:LIM3:DEL
Display Lines
Opens a submenu to enable, disable and set display lines. Which softkeys are available
depends on the display mode (frequency or time range). For details see also "Display Lines" on
page 4.122 and "To work with display lines" on page 4.119.
Display Line 1 and Display Line 2
Enable or disable the level lines 1/2 and open an edit dialog box to enter the position of the
lines. For details see also "Display Lines" on page 4.122 and "To work with display lines" on
page 4.119.
Remote: CALC:DLIN:STAT ON
Remote: CALC:DLIN –20dBm
Frequency Line 1 and Frequency Line 2 (span > 0)
Enable or disable the frequency lines 1/2 and open an edit dialog box to enter the position of the
lines. For details see also "Display Lines" on page 4.122 and "To work with display lines" on
page 4.119.
Remote: CALC:FLIN:STAT ON
Remote: CALC:FLIN 120MHz