R&S FSL Analog Demodulation (Option K7)
1300.2519.12 4.157 E-11
In Analog Demodulation mode, the next measurement is triggered if the selected input signal
exceeds the threshold specified using the Trg / Gate Level softkey. A periodic signal modulated
onto the carrier frequency can be displayed in this way. It is recommended that the measurement
time covers at least five periods of the audio signal.
For further details refer to the Trg / Gate Source softkey in the trigger menu.
Remote: TRIG:SOUR IMM | IFP | EXT | FM | PM | AM | RF (Free Run, IF Power,
E
xtern
, FM (Offline), PM (Offline), AM (Offline), RF (Offline))
Trigger Level
For details refer to the Trg / Gate Level softkey in the in the trigger menu of the base unit.
Trigger Polarity Pos/Neg
For details refer to the Trg / Gate Polarity Pos/Neg softkey in the in the trigger menu of the
base unit.
Trigger Offset
For details on the relation between demodulation bandwidth (option Analog Demodulation,
R&S FSL–K7) and trigger offset refer to "Sample rate, measurement time and trigger offset" on
page 4.143.
For details refer to the Trigger Offset softkey in the in the trigger menu of the base unit.