R&S FSL Triggering the Sweep – TRIG Key
1300.2519.12 4.33 E-11
Softkeys of the trigger menu
The following table shows all softkeys available in the trigger menu. It is possible that your instrument
configuration does not provide all softkeys. If a softkey is only available with a special option, model or
(measurement) mode, this information is delivered in the corresponding softkey description.
Menu / Command Command
Trg / Gate Source
Trg / Gate Level
Trg / Gate Polarity Pos/Neg
Trigger Offset / Repetition Intervall
Gated Trigger
Gate Settings Gate Mode Lvl/Edge
Gate Delay
Gate Length
Trg / Gate Source
Trg / Gate Level
Trg / Gate Polarity Pos/Neg
Sweep Time
More
IF Power Retrigger Holdoff
IF Power Retrigger Hysteresis
TV Trig Settings Vert Sync
Vert Sync Odd Field
Vert Sync Even Field
Hor Sync
Video Pol Pos/Neg
Lines 625/525
TV Free Run On/Off
Trg / Gate Source
Opens the Trigger / Gate Source dialog box to select the trigger / gate mode. For detailed
information on trigger modes see "Trigger mode overview" on page 4.31.
The gate–related settings are only available with option Gated Sweep, R&S FSL–B8. As gate
modes, all modes apart from the TV Trigger mode (option TV Trigger, B6, available from
firmware version 1.10) are available. For details see also "To use gated sweep operation (option
Gated Sweep, B8)" on page 4.29.
The default setting is Free Run. If a trigger mode other than Free Run has been set, the
enhancement label TRG is displayed.
Remote: TRIG:SOUR IMM | VID | IFP | EXT | TV | TIME (Free Run, Video, IF Power,
Extern
, TV Trigger, Time Trigger)
Remote: SWE:EGAT:SOUR EXT (Extern)