EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 356

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Bluetooth Measurements (Option K8)
1300.2519.12 4.171 E-11
Search Len Manual
Opens an edit dialog box to enter the search length used for determining the sync word or the
burst. The unit of the search length is seconds; the values range from 100 µs to (400001 / points
per symbol) µs. The default setting is 1825 µs.
The selected number of points / symbol and the maximum search length is listed in the table
below.
Points per symbol Max. search length
(number of slots)
2 104.4
4 52.2
8 26.1
16 13.1
32 6.5
For information on the correlation of trigger and record length refer to the Search Len Auto
softkey.
Remote: DDEM:SEAR:TIME 100US
Select Trace
Opens an edit dialog box to select the measurement curve whose numeric results will be
displayed in the lower half of the screen. The default curve selected is trace 1.
Remote: CONF:BTO:TRAC2:SEL
Points / Symbol
(Output Power, TX Spec ACP, Modulation Char, Init Carr Freq Tol, Carr Freq Drift)
Opens an edit dialog box to change the number of measurement samples per symbol. For Basic
Rate measurements, the possible values are 2, 4, 8, 16, 32. The default value is 4. For EDR
measurements, the default value is set and cannot be changed.
The RF Test Specification requests an oversampling factor of at least 4. With this oversampling
factor a 5 Slot Packet corresponds to 12500 measurement samples (= 2500 samples / slot).
Remote: CONF:BTO:PRAT 16

Table of Contents