EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 361

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Bluetooth Measurements (Option K8) R&S FSL
1300.2519.12 4.176 E-11
Res BW Manual (Output Power, Rel TX Power)
Opens an edit dialog box to enter the resolution bandwidth. The values range from 1 MHz to 3
MHz. The default value is 3 MHz.
Remote: CONF:BTO:BAND 1KHZ
Res BW Auto (Output Power, TX Spec ACP, Rel TX Power, Spurious Emissions)
Sets the bandwidth according to the values defined in the RF Test Specification.
Remote: CONF:BTO:BAND:AUTO ON
Video BW Manual (TX Spec ACP)
The default value is 300 kHz, according to the values defined in the RF Test Specification.
For further details refer to the Video BW Manual softkey in the bandwidth menu of the base
unit.
Remote: CONF:BTO:BAND:VID 100HZ
Video BW Auto (TX Spec ACP, Spurious Emissions)
Sets the video bandwidth according to the values defined in the RF Test Specification. In the In–
band Spurious Emissions measurement, the video bandwidth is always adjusted automatically.
Therefore the softkey is activated by default and its state cannot be changed.
Remote: CONF:BTO:BAND:VID:AUTO ON
Filter Type (Output Power, TX Spec ACP, Spurious Emissions, Rel TX Power)
Opens the Filter Type dialog box to select the filter type. For the measurements Output Power
and Rel TX Power, a Gaussian or a channel filter can be selected. For the measurements TX
Spec ACP and Spurious Emissions, the filter type is a Gaussian filter.
Remote: BAND:TYPE NORM
Meas Filter (Modulation Char, Init Carr Freq Tol, Carr Freq Drift)
Activates or deactivates a filter that limits the bandwidth for the modulation measurements.
Since the RF Test Specification Rev 2.0.E.3 it is required to use this filter and therefore it is
activated by default.
The filter is flat within 1.04 MHz (ripple: only 0.02 dB) and has steep slopes outside this area.
The filter has the following characteristics:
passband ripple up to
±
550 kHz < 0.5 dB (peak to peak)
minimum attenuation in the transition band
±
650 kHz:
3 dB
±
1 MHz:
14 dB
±
2 MHz:
44 dB
Remote: DDEM:FILT:MEAS BTO