Spectrogram Measurement (Option K14) R&S FSL
1300.2519.12 4.198 E-11
Select 1 2 3 4
The edit dialog box for markers and delta markers is extended as described in "Markers and
marker values" on page 4.193.
For further details refer to the Select 1 2 3 4
softkey in the marker–> menu of the base unit.
Remote: CALC:MARK2:SPEC:FRAM –20
Remote: CALC:DELT3:SPEC:FRAM –50
Peak
Sets the active marker/delta marker to the highest maximum according to the selected search
type (Marker Search Type softkey).
Remote: CALC:MARK:MAX (in x direction)
Remote: CALC:DELT:MAX (in x direction)
Remote: CALC:MARK:SPEC:Y:MAX (y direction)
Remote: CALC:DELT4:SPEC:Y:MAX (y direction)
Remote: CALC:MARK2:SPEC:XY:MAX (xy direction)
Remote: CALC:DELT2:SPEC:XY:MAX (xy direction)
Next Peak
Sets the active marker/delta marker to the next maximum according to the mode selected using
the Search Mode softkey. For the XY Search (Marker Search Type softkey), this softkey is not
available.
Remote: for search in x direction see Next Peak X Search < abs > softkey
Remote: for search in y direction see Next Peak Y Search up/abs/dn softkey
Search Mode
Displays a submenu to define the search mode for the Next Peak and Next Min softkeys.
Next Peak X Search < abs >
Selects the mode of the Next Peak softkey, if the X Search is selected (Marker Search Type
softkey). Three settings are available:
< Sets the active marker/delta marker to the next maximum left to the marker of the
selected frame.
abs Sets the active marker/delta marker to the next lower maximum of the selected
frame.
> Sets the active marker/delta marker to the next maximum right to the marker of the
selected frame.