Noise Figure Measurements Option (K30) R&S FSL
1300.2519.12 4.254 E-11
Start Freq
Specifies the start frequency. This is the first receive frequency (RF) entry in the Frequency
Table and the Frequency List Results (result display).
If the start frequency is changed, the Frequency Table is updated accordingly.
Remote: FREQ:STAR 500MHZ
Stop Freq
Specifies the stop frequency. This is the last receive frequency (RF) entry in the Frequency
Table and the Frequency List Results (result display).
If the stop frequency is changed, the Frequency Table is updated accordingly.
Remote: FREQ:STOP 700MHZ
Step Freq
Specifies the step size between the single measurement steps. If the step frequency is larger
than the difference between the start frequency and the stop frequency, the Frequency Table
and the Frequency List Results (result display) just contain the start and stop frequency.
If the step frequency is changed, the Frequency Table is updated accordingly.
Remote: FREQ:STEP 10MHZ
Mode
Specifies the measurement mode. For details on modes refer to "Measurement modes of the
noise figure measurement" on page 4.248.
If the mode is changed, the Frequency Table is updated accordingly.
Remote: SENS:CONF:MODE:DUT DOWN
Fixed LO
Specifies the fixed local oscillator frequency. This field is only available if a frequency–converting
measurement mode is selected (Mode field). For details on modes refer to Measurement modes
of the noise figure measurement.
If the fixed LO is changed, the Frequency Table is updated accordingly.
Remote: SENS:CONF:MODE:SYST:LOSC:FREQ 1MHZ