Image Rej
Specifies the suppression applied to the second sideband. This field is only available if a
frequency–converting measurement mode is selected. For details on modes refer to
"Measurement modes of the noise figure measurement" on page 4.248.
The value entered is applied across the complete frequency range. A value of 999.99 dB
corresponds to the generally used single–sideband measurement (SSB), where the second
sideband does not noticeably affect the measurement result. This is the default value. A value of
0 dB corresponds to the double–sideband measurement (DSB), where both sidebands are
converted to the same extent.
Remote: CORR:IREJ 100
Display List/Graph
Configures the result display. The measurement results are displayed either in form of a list of
measurement points or as a graphical trace. For further details refer to "Result displays" on page
4.250.
Remote: DISP:TABL ON
Display Settings
Opens the Graphic dialog box to modify the graphical results display.
Under Results Settings, the settings that affect the overall results display are defined. Under
Noise Trace Settings, the settings related to the graphical display of noise results are defined.
Under Gain Trace Settings, the settings related to the graphical display of gain results are
defined.
The Graphic dialog box contains the following elements:
Results Settings Combined Trace Display
Noise Trace Settings Y–Axis
Automatic Scaling
Min Y–Axis NF
Min Y–Axis Temp
Max Y–Axis NF