CDMA2000 BTS Analyzer (Option K82) R&S FSL
1300.2519.12 4.320 E-11
The Result Settings dialog box contains the following parameters:
Power Control Group
Channel (Code) Number
Code Order
Code Power
Power Reference
Normalize
Power Control Group (Result Settings dialog box)
In some measurements it is possible to highlight the data of a specific PCG (Power Control
Group). Select the PCG (Power Control Group) on which to put the focus on in this field. The
range of the value depends on the Capture Length defined in the IQ Capture Settings dialog
box. The range is (0 to Capture Length-1).
The defined value is valid for any measurement which takes PCGs into account (for further
information see also the Select Ch/PCG softkey). The selected PCG is highlighted in red on the
screen (as opposed to the others which are displayed in yellow).
Remote: CDP:SLOT 0 ...(IQ_CAPTURE_LENGTH–1)
Channel (Code) Number (Result Settings dialog box)
The entry in this field corresponds to a specific code to be examined. The valid range is from 0
to BSF-1. The base spreading factor (BSF) is either 64 or 128 depending on the setting of the
Base SF field and can not be edited via this field. The selected channel will be marked red in the
Channel Table. In the Code Domain Power display and the Code Domain Error Power display,
all codes belonging to this channel will be marked red.
The defined value is valid for any measurement that takes a selected channel into account (for
further information see also the Select Ch/PCG softkey).
The rotary knob behavior depends on the result display and the graphic display. In the case of
the Code Domain Power and Code Domain Error result displays, the rotary knob behavior
depends on the code order (refer to Code Order) The adjacent channel is always selected with
the rotary knob. In the channel table, the rotary knob is used to scroll through the list.
Remote: CDP:CODE 0...(BASE SF–1)