EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 682

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Instrument Setup and Interface Configuration - SETUP Key
1300.2519.12 4.497 E-11
Delete
Deletes the selected factor after confirmation.
This softkey is available from firmware version 1.10.
Remote: CORR:TRAN DEL
Ref Level Adjust Man/Auto
Activates or deactivates the automatic adjustment of the reference level to the selected
transducer factor.
Auto
Activates the automatic adjustment.
Man
Deactivates the automatic adjustment. Adjust the reference level via the amplitude
menu.
If a transducer factor is used (active), the trace is moved by a calculated shift. However, an
upward shift reduces the dynamic range for the displayed values. With this softkey set to Auto,
the original dynamic range is restored by also shifting the reference level by the maximum value
of the transducer factor.
This softkey is available from firmware version 1.10.
Remote: CORR:TRAN:ADJ:RLEV ON
Show Directory
Displays the subdirectory of the selected directory.
This softkey is available from firmware version 1.10.
IF Output (option Additional Interfaces, B5 only)
Opens the IF Output Source dialog box to select the output of the IF/VIDEO OUT connector: IF,
Video 200 mV, or Video 1 V.
Note: The AF output available at the frontpanel can only be used if the IF output source is set to
video.
For further details refer to the Quick Start Guide, chapter 1 "Front and Rear Panel".
Remote: OUTP IF VID
Noise Src On/Off (option Additional Interfaces, B5 only)
Switches the supply voltage for an external noise source on or off. For details on connectors
refer to the Quick Start Guide, chapter 1 "Front and Rear Panel" .
Remote: DIAG:SERV:NSO ON