Instrument Setup and Interface Configuration - SETUP Key R&S FSL
1300.2519.12 4.498 E-11
Alignment
Opens a submenu with the available functions for recording, displaying and activating the data
for self alignment.
The correction data and characteristics required for the alignment are determined by comparison
of the results at different settings with the known characteristics of the high-precision calibration
signal source at 65.83 MHz. The correction data are stored as a file on flash disk and can be
displayed using the Show Align Results softkey.
Self Align
Starts the recording of correction data of the instrument. If the correction data acquisition has
failed or if the correction values are deactivated, a corresponding message is displayed in the
status field.
As long as the self alignment data is collected the procedure can be cancelled using the Abort
button.
Remote: *CAL?
Show Align Results
Opens a dialog box that displays the correction data of the alignment:
– date and time of last correction data record
– overall results of correction data record
– list of found correction values according to function/module
The results are classified as follows:
PASSED
calibration successful without any restrictions
CHECK
deviation of correction value larger than expected, correction could
however be performed
FAILED
deviations of correction value too large, no correction was possible. The
found correction data are not applicable.
As long as the self alignment data is collected the procedure can be cancelled using the Abort
button.
Remote: CAL:RES?
Remote: CAL:ABOR
Corr Data On/Off
Activates and deactivates the alignment data, e.g. for service purposes the correction data can
be deactivated.
Remote: CAL:STAT ON
General Setup
Opens a submenu for all general settings such as IP address and LAN settings, date and time,
remote control (optional) and MEAS display.