R&S FSL Instrument Setup and Interface Configuration - SETUP Key
1300.2519.12 4.499 E-11
Configure Network
Opens the Network Connections dialog box to change the LAN settings. For details refer to the
Quick Start Guide, chapter 2 "Preparing for Use" and appendix B "LAN Interface".
Network Address
Opens a submenu to configure the internet protocol properties and the computer name.
Computer Name
Opens an edit dialog box to enter the computer name via the keypad. The naming conventions
of Windows apply. If too many characters and/or numbers are entered, in the status line, an
according message is displayed. For step-by-step instructions refer to the Quick Start Guide,
appendix B "LAN Interface".
IP Address
Opens an edit dialog box to enter the IP address via the keypad. The TCP/IP protocol is
preinstalled with the IP address 10.0.0.10. If the DHCP server is available (DHCP On), the
dialog box entry is read-only.
The IP address consists of four number blocks separated by dots. Each block contains 3
numbers in maximum (e.g. 100.100.100.100), but also one or two numbers are allowed in a
block (as an example see the preinstalled address). For step-by-step instructions refer to the
Quick Start Guide, chapter 2 "Preparing for Use".
Subnet Mask
Opens an edit dialog box to enter the subnet mask via the keypad. The TCP/IP protocol is
preinstalled with the subnet mask 255.255.255.0. If the DHCP server is available (DHCP On),
the dialog box entry is read-only.
The subnet mask consists of four number blocks separated by dots. Each block contains 3
numbers in maximum (e.g. 100.100.100.100), but also one or two numbers are allowed in a
block (as an example see the preinstalled address). For step-by-step instructions refer to the
Quick Start Guide, chapter 2 "Preparing for Use".
DHCP On/Off
Switches between DHCP server available (On) or not available (Off). If a DHCP server is
available in the network, the IP address and subnet mask of the instrument are obtained
automatically from the DHCP server. For further details refer to the Quick Start Guide, chapter 2
"Preparing for Use".