R&S FSL CALCulate Subsystem
1300.2519.12 6.13 E-11
Characteristics
RST value: OFF
SCPI: device–specific
Mode
all
CALCulate<1|2>:DELTamarker<1...4>:AOFF
This command switches off all active delta markers.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
Example
CALC:DELT:AOFF
Switches off all delta markers.
Characteristics
RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, WCDMA
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe]
This command switches the relative measurement to a fixed reference value on or off. Marker 1
will be activated previously and a peak search will be performed, if necessary. If marker 1 is
activated, its position becomes the reference point for the measurement. The reference point
can then be modified with the
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:X commands and
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y independently of
the position of marker 1 and of a trace. It applies to all delta markers as long as the function is
active.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
Example
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
A