CALCulate Subsystem R&S FSL
1300.2519.12 6.14 E-11
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
This command sets the reference point level for all delta markers for a measurement with fixed
reference point (CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe]) to
the peak of the selected trace.
For phase–noise measurements
(CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:PNOise[:STATe]), the command
defines a new reference point level for delta marker 2.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
Parameter
<numeric_value>
Example
CALC:DELT:FUNC:FIX:RPO:MAX
Sets the reference point level for delta markers to the peak of the selected trace.
Characteristics
RST value: –
SCPI: device–specific
Mode
A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:X
This command defines a new frequency reference (span > 0) or time (span = 0) for all delta
markers for a measurement with fixed reference value
(CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe]).
For phase–noise measurements
(CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:PNOise[:STATe]), the command
defines a new frequency reference or time for delta marker 2.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
<numeric_value>
Example
CALC:DELT:FUNC:FIX:RPO:X 128MHz
Sets the frequency reference to 128 MHz.
Characteristics
RST value: – (CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe] is
set to OFF)
SCPI: device–specific
Mode
A