R&S FSL CALCulate Subsystem
1300.2519.12 6.21 E-11
C
ALCulate<1|2>:DELTamarker<1...4>:X
This command positions the selected delta marker to the indicated frequency (span > 0), time
(span = 0) or level (APD measurement = ON or CCDF measurement = ON). The input is in
absolute values or relative to marker 1 depending on the command
CALCulate<1|2>:DELTamarker<1...4>:MODE. If reference fixed measurement
(CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe] is ON) is active,
relative values refer to the reference position are entered. The query always returns absolute
values.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
0 to MAX (frequency | sweep time)
Example
CALC:DELT:MOD REL
Switches the input for all delta markers to relative to marker 1.
CALC:DELT2:X 10.7MHz
Positions delta marker 2 10.7 MHz to the right of marker 1.
CALC:DELT:X?
Outputs the absolute frequency/time of delta marker 1.
CALC:DELT:X:REL?
Outputs the relative frequency/time/level of delta marker 1.
Characteristics
RST value:
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, WCDMA
CALCulate<1|2>:DELTamarker<1...4>:X:RELative?
This command queries the frequency (span > 0) or time (span = 0) of the selected delta marker
relative to marker 1 or to the reference position ( for
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe] is ON). The
command activates the corresponding delta marker, if necessary.
The numeric suffixes <1|2> are irrelevant for this command.
Example
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or relative to the reference position.
Characteristics
RST value:
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, WCDMA