CALCulate Subsystem R&S FSL
1300.2519.12 6.32 E-11
Example
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Characteristics
*RST value: OFF
SCPI: conform
Mode
A, BT, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:TRACe
This command assigns a limit line to a trace.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
1 to 6
Example
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
Characteristics
*RST value: 1
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:UNIT
This command defines the unit of the selected limit line.
Upon selection of the unit DB the limit line is automatically switched to the relative mode. For
units different from DB the limit line is automatically switched to absolute mode.
The units DEG, RAD, S, HZ, PCT are not available in spectrum analyzer mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
DBM | DBPW | WATT | DBUV | DBMV | VOLT |DBUA | AMPere | DB | DEG | RAD | HZ | PCT
Example
CALC:LIM4:UNIT DBUV
Sets the unit of limit line 4 to dBµV.
Characteristics
*RST value: DBM
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO