CALCulate Subsystem R&S FSL
1300.2519.12 6.34 E-11
The numeric suffixes <1|2> and <1...8> are irrelevant for this command.
Parameter
first value: 0 to 100dB; limit for the upper (lower) adjacent channel
second value: 0 to 100dB; is ignored but must be indicated for reasons of compatibility with the
F
SP family
Example
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper adjacent channel to 30 dB
below the channel power.
Characteristics
*RST value: 0 dB
SCPI: device–specific
Mode
A, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:ACPower:ACHannel[:RELative]:STATe
This command activates the limit check for the relative limit value of the adjacent channel when
adjacent–channel power measurement is performed. Before the command, the limit check must
be activated using CALCulate<1|2>:LIMit<1...8>:ACPower[:STATe].
The result can be queried with
CALCulate<1|2>:LIMit<1...8>:ACPower:ACHannel:RESult?. It should be noted that a
complete measurement must be performed between switching on the limit check and the result
query, since otherwise no correct results are available.
The numeric suffixes <1|2> and <1...8> are irrelevant for this command.
Parameter
ON | OFF
Example
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper adjacent channel to 30 dB
below the channel power.
CALC:LIM:ACP:ACH:ABS –35DBM, –35DBM
Sets the absolute limit value for the power in the lower and upper adjacent channel to –35 dBm.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent channel measurement.
CALC:LIM:ACP:ACH:STAT ON
Switches on the check of the relative limit values for adjacent channels.
CALC:LIM:ACP:ACH:ABS:STAT ON
Switches on the check of absolute limit values for the adjacent channels.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A, CDMA, EVDO

Table of Contents