EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 804

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL CALCulate Subsystem
1300.2519.12 6.45 E-11
CALCulate<1|2>:LIMit<1…8>:ESPectrum:PCLass<1…4>:MAXimum
This command sets the upper limit level for one power class. The unit is dBm. The limit always
ends at + 200 dBm, i.e. the upper limit of the last power class can not be set. If more than one
power class is in use, the upper limit must equal the lower limit of the next power class.
The numeric suffix at PCLass<1…4> denotes the power class to be defined.
The numeric suffixes <1|2> and <1…8> are irrelevant for this command.
Parameter
<numeric_value>
Example
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40 dBm.
Characteristics:
*RST value: –
SCPI: conform
Mode
A, CDMA, EVDO
CALCulate<1|2>:LIMit<1…8>:ESPectrum:PCLass<1…4>:COUNt
This command sets the number of power classes to be defined.
The numeric suffixes are irrelevant
Parameter
1 to 4
Example
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Characteristics:
*RST value: –
SCPI: conform
Mode
A, CDMA, EVDO
CALCulate<1|2>:LIMit<1…8>:ESPectrum:PCLass<1…4>:LIMit[:STATe]
This command defines which limits are evaluated in the measurement.
The numeric suffix at PCLass<1..4> defines the power class to be evaluated. The numeric
suffixes <1|2> and <1..8> are irrelevant
Parameter
ABSolute Evaluates only limit lines with absolute power values
RELative Evaluates only limit lines with relative power values
OR Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
AND Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.

Table of Contents

Related product manuals