EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 808

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL CALCulate Subsystem
1300.2519.12 6.49 E-11
C
ALCulate<1|2>:LIMit<1...8>:LOWer:MARGin
This command defines a margin to a lower limit line, at which out–of–limit values are signaled (if
the limit check is active), but not handled as a violation of the limit value.
Only the unit dB is available in spectrum analyzer mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
<numeric_value>
Example
CALC:LIM:LOW:MARG 10dB
Characteristics
*RST value: 0
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:LOWer:MODE
This command selects the relative or absolute scaling for the y–axis of the selected lower limit
line.
Selecting RELative causes the unit to be switched to DB.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
RELative | ABSolute
Example
CALC:LIM:LOW:MODE REL
Defines the y–axis of limit line 2 as relative scaled.
Characteristics
*RST value: ABSolute
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:LOWer:SHIFt
This command shifts a limit line by the indicated value in Y direction. In contrast to
CALCulate<1|2>:LIMit<1...8>:LOWer:OFFSet, the line is shifted by modifying the
individual Y values but not by means of an additive offset.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.10.
Parameter