CALCulate Subsystem R&S FSL
1300.2519.12 6.50 E-11
Example
CALC:LIM3:LOW:SHIF 20DB
Shifts all Y values of limit line 3 by 20 dB.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:LOWer:SPACing
This command selects linear or logarithmic interpolation for the lower limit line.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
LINear | LOGarithmic
Example
CALC:LIM:LOW:SPAC LIN
Characteristics
*RST value: LIN
SCPI: device–specific
Mode
A, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:LOWer:THReshold
This command defines an absolute threshold value for limit lines with relative y–axis scaling.
The absolute threshold value is used in the limit check as soon as it exceeds the relative limit
value.
The unit must correspond to the unit selected with CALCulate<1|2>:LIMit<1...8>:UNIT
(except dB which is not allowed). If no unit is indicated, the unit defined with
CALCulate<1|2>:LIMit<1...8>:UNIT is automatically used (exception: dBm instead of
dB).
The units DEG, RAD, S, HZ, PCT are not available in the spectrum analyzer mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
<numeric_value>
Example
CALC:LIM2:LOW:THR –35DBM
Defines an absolute threshold value for limit line 2.
Characteristics
*RST value: –200 dBm
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO