R&S FSL CALCulate Subsystem
1300.2519.12 6.51 E-11
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line. This subsystem is available from
firmware version 1.10.
Commands of the CALCulate:LIMit:UPPer Subsystem
– CALCulate<1|2>:LIMit<1...8>:UPPer[:DATA]
– CALCulate<1|2>:LIMit<1...8>:UPPer:MARGin
– CALCulate<1|2>:LIMit<1...8>:UPPer:MODE
– CALCulate<1|2>:LIMit<1...8>:UPPer:OFFSet
– CALCulate<1|2>:LIMit<1...8>:UPPer:SHIFt
– CALCulate<1|2>:LIMit<1...8>:UPPer:SPACing
– CALCulate<1|2>:LIMit<1...8>:UPPer:STATe
– CALCulate<1|2>:LIMit<1...8>:UPPer:THReshold
CALCulate<1|2>:LIMit<1...8>:UPPer[:DATA]
This command defines the values for the upper limit lines
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer
limit line has to be identical. Otherwise default values are entered for missing values or not
necessary values are deleted.
The unit must be identical with the unit selected by CALCulate<1|2>:LIMit<1...8>:UNIT.
If no unit is indicated, the unit defined with CALCulate<1|2>:LIMit<1...8>:UNIT is
automatically used.
In spectrum analyzer mode, the limit check indicates errors if the measured values exceed the
UPPer limit line. The units DEG, RAD, S, HZ, PCT are not available in spectrum analyzer mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
<numeric_value>,<numeric_value>
Example
CALC:LIM2:UPP –10,0,0,–10,–5
Defines 5 upper limit values for limit line 2 in the preset unit.
CALC:LIM2:UPP?
Outputs the upper limit values for limit line 2 separated by a comma.
Characteristics
*RST value: – (CALCulate<1|2>:LIMit<1...8>:STATe is set to OFF)
SCPI: conform
Mode
A, ADEMOD, CDMA, EVDO, NF