R&S FSL CALCulate Subsystem
1300.2519.12 6.53 E-11
Parameter
<numeric_value>
Example
CALC:LIM2:UPP:OFFS 3dB
Shifts limit line 2 by 3 dB upwards.
Characteristics
*RST value: 0
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:UPPer:SHIFt
This command moves a limit line by the indicated value in Y direction. In contrast to
CALCulate<1|2>:LIMit<1...8>:UPPer:OFFSet, the line is shifted by modifying the
individual Y values and not by means of an additive offset.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.10.
Parameter
<numeric_value>
Example
CALC:LIM3:UPP:SHIF 20
Shifts all Y values of limit line 3 by 20 limit line units, e.g. dB.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:UPPer:SPACing
This command selects linear or logarithmic interpolation for the upper limit line.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
LINear | LOGarithmic
Example
CALC:LIM:UPP:SPAC LIN
Characteristics
*RST value: LIN
SCPI: device–specific
Mode
A, CDMA, EVDO