CALCulate Subsystem R&S FSL
1300.2519.12 6.56 E-11
C
ALCulate<1|2>:MARKer<1...4>[:STATe]
This command switches on or off the currently selected marker. If no indication is made, marker
1 is selected automatically. If marker 2, 3 or 4 is selected and used as a delta marker, it is
switched to marker mode.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
Parameter
ON | OFF
Example
CALC:MARK3 ON
Switches on marker 3 or switches to marker mode.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:AOFF
This command switches off all active markers and all delta markers and active marker/delta
marker measurement functions.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
CALC:MARK:AOFF
Switches off all markers.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, WCDMA
CALCulate<1|2>:MARKer<1...4>:COUNt
This command switches on or off the frequency counter at the marker position.
The count result is queried with CALCulate<1|2>:MARKer<1...4>:COUNt:FREQuency?.
Frequency counting is possible only for one marker at a time. If it is activated for another
marker, it is automatically deactivated for the previous marker.
It should be noted that a complete sweep must be performed after switching on the frequency
counter to ensure that the frequency to be measured is actually reached. The synchronization to
the sweep end required for this is only possible in single sweep mode.