R&S FSL CALCulate Subsystem
1300.2519.12 6.57 E-11
Parameter
ON | OFF
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A
CALCulate<1|2>:MARKer<1...4>:COUNt:FREQuency?
This command queries the result of the frequency counter for the indicated marker. Before the
command, the frequency counter should be switched on and a complete measurement
performed to obtain a correct count result. Therefore, a single sweep with synchronization must
be performed between switching on the frequency counter and querying the count result.
The numeric suffixes <1|2> are irrelevant for this command.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK2:COUN ON
Switches the frequency counter for marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:COUN:FREQ?
Outputs the measured value of marker 2.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A