CALCulate Subsystem R&S FSL
1300.2519.12 6.58 E-11
C
ALCulate<1|2>:MARKer<1...4>:COUNt:RESolution
This command specifies the resolution of the frequency counter.
T
he numeric suffixes <1|2> and <1...4> are irrelevant for this command.
Parameter
0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
Example
CALC:MARK:COUN:RES 1kHz
Sets the resolution of the frequency counter to 1 kHz.
Characteristics
*RST value: 1kHz
SCPI: device–specific
Mode
A
CALCulate<1|2>:MARKer<1...4>:LOEXclude
This command switches the local oscillator suppression for peak search on or off. This setting
applies to all markers and delta markers.
The numeric suffixes <1|2> and <1…4> are irrelevant.
Parameter
ON | OFF
Example
CALC:MARK:LOEX ON
Characteristics
*RST value: ON
SCPI: device–specific
Mode
A–F, ADEMOD, CATV, SPECM
CALCulate<1|2>:MARKer<1...4>:MAXimum[:PEAK]
This command positions the marker to the current maximum value of the corresponding trace.
The corresponding marker is activated first or switched to the marker mode.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no maximum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:MAX
Positions marker 2 to the maximum value of the trace.