CALCulate Subsystem R&S FSL
1300.2519.12 6.64 E-11
C
ALCulate<1|2>:MARKer<1...4>:X
This command positions the selected marker to the indicated frequency (span > 0), time (span =
0) or level (APD measurement or CCDF measurement ON).
If marker 2, 3 or 4 is selected and used as delta marker, it is switched to marker mode.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
Parameter
0 to MAX (frequency | sweep time)
Example
CALC:MARK2:X 10.7MHz
Positions marker 2 to frequency 10.7 MHz.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:X:SLIMits[:STATe]
This command switches between a limited (ON) and unlimited (OFF) search range.
If the power measurement in zero span is active, this command limits the evaluation range on
the trace.
The numeric suffixes <1|2> and <1...4> are irrelevant for this command.
Parameter
ON | OFF
Example
CALC:MARK:X:SLIM ON
Switches on search limitation.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A, ADEMOD, CATV, SPECM