CALCulate Subsystem R&S FSL
1300.2519.12 6.84 E-11
C
ALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:LIST?
This command reads out the list of the harmonics. The first value is the absolute power of the
first harmonic (the unit is set via the CALCulate<1|2>:UNIT:POWer command). The following
values are relative to the carrier signal and have the unit dB. They are separated by commas
and correspond to the harmonics to be measured (set via the
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:NHARmonics command).
To obtain a correct result, a complete sweep with synchronization to the end of the sweep must
be performed before a query is output. Synchronization is possible only in the single sweep
mode.
The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.10.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:HARM:NHARM 3
Sets the number of harmonics to be measured to 3.
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
INIT;*WAI
Starts a sweep and waits for the end of the sweep.
CALC:MARK:FUNC:HARM:LIST?
Returns the values for the 3 measured harmonics.
Characteristics
RST value: –
SCPI: device–specific
Mode
A–F, A–T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:NHARmonics
This command sets the number of harmonics to be measured.
The numeric suffix <1|2> and <1...4> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
1...26
Example
CALC:MARK:FUNC:HARM:NHARM 3
Sets the number of harmonics to be measured to 3.
Characteristics
RST value: 10
SCPI: device–specific
Mode
A–F, A–T