R&S FSL CALCulate Subsystem
1300.2519.12 6.99 E-11
CALC:MARK:FUNC:SUMM:MEAN:AVER:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
This command queries the result of the measurement of the mean value with active peak hold.
The query is only possible if the peak hold function has been switched on previously using
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PHOLd.
The query is possible only if the peak hold function is active.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:MEAN ON
Switches on the function.
CALC:MARK:FUNC:SUMM:PHOL ON
Switches on the peak value measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:MEAN:PHOL:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:MEAN:RESult?
This command queries the result of the measurement of the mean value.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.