CALCulate Subsystem R&S FSL
1300.2519.12 6.100 E-11
CALC:MARK:FUNC:SUMM:MEAN ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:MEAN:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:MODE
This command selects absolute or relative power measurement in zero span.
The reference power for relative measurement is defined with
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:REFerence:AUTO. If the reference power
is not defined, the value 0 dBm is used.
The numeric suffixes <1|2> and <1...4> are irrelevant for this command.
Parameter
ABSolute | RELative
Example
CALC:MARK:FUNC:SUMM:MODE REL
Switches the power measurement in zero span to relative.
Characteristics
RST value: ABSolute
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PHOLd
This command switches on or off the peak–hold function for the active power measurement in
zero span.
The peak–hold function is reset by switching it off and on again.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
CALC:MARK:FUNC:SUMM:PHOL ON
Switches on the function.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A–T