R&S FSL CALCulate Subsystem
1300.2519.12 6.103 E-11
C
haracteristics
*RST value: –
SCPI: device–specific
M
ode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:REFerence:AUTO
This command sets the currently measured average value
(CALCulate<1|2>:MARKer:FUNCtion:SUMMary:MEAN[:STATe]) and RMS value
(CALCulate<1|2>:MARKer:FUNCtion:SUMMary:RMS[:STATe]) as reference values for
relative measurements in zero span.
If the measurement of RMS value and average is not activated, the reference value 0 dBm is
used.
If the function CALCulate<1|2>:MARKer:FUNCtion:SUMMary:AVERage or
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PHOLd is switched on, the current value is
the accumulated measurement value at the time considered.
The numeric suffixes <1|2> and <1...4> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
Parameter
ONCE
Example
CALC:MARK:FUNC:SUMM:REF:AUTO ONCE
Takes the currently measured power as reference value for the relative power measurement in
zero span.
Characteristics
RST value:
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:RMS[:STATe]
This command switches on or off the measurement of the effective (RMS) power. If necessary
the function is switched on previously.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
CALC:MARK:FUNC:SUM:RMS ON
Switches on the function.
Characteristics
*RST value: OFF
SCPI: device–specific