CALCulate Subsystem R&S FSL
1300.2519.12 6.102 E-11
C
ALCulate<1|2>:MARKer:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command is used to query the result of the measurement of the positive peak value with
active peak hold function. The query is only possible if the peak hold function has been
activated previously using CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PHOLd.
A
complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
CALC:MARK:FUNC:SUMM:PHOL ON
Switches on the measurement of the peak value.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:PPE:PHOL:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PPEak:RESult?
This command is used to query the result of the measurement of the positive peak value. The
measurement may have to be switched on previously.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:PPE:RES?
Outputs the result.