R&S FSL CALCulate Subsystem
1300.2519.12 6.111 E-11
CALCulate<1|2>:PSEarch|PEAKsearch:SUBRanges
This command sets the number of peaks per range that are stored in the list. Once the selected
number of peaks has been reached, the peak search is stopped in the current range and
continued in the next range.
The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
1 to 50
Example
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Characteristics
RST value: 25