CALCulate Subsystem R&S FSL
1300.2519.12 6.110 E-11
CALCulate:PSEarch|PEAKsearch Subsystem
The CALCulate:PSEarch|PEAKsearch subsystem contains the remote commands for Spurious
Emissions measurements. Both groups of commands (PSEarch and PEAKsearch) perform the same
functions.
Commands of the CALCulate:PSEarch|PEAKsearch Subsystem
CALCulate<1|2>:PSEarch|PEAKsearch[:IMMediate]
CALCulate<1|2>:PSEarch|PEAKsearch:AUTO
CALCulate<1|2>:PSEarch|PEAKsearch:MARGin
CALCulate<1|2>:PSEarch|PEAKsearch:PSHow
CALCulate<1|2>:PSEarch|PEAKsearch:SUBRanges
CALCulate<1|2>:PSEarch|PEAKsearch[:IMMediate]
This command determines the list of the subrange maximums from the existing sweep results.
The numeric suffixes <1|2> are not relevant.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.80.
Example
CALC:PSE
Starts to determine the list.
Characteristics
RST value:
SCPI: device–specific
Mode
A
CALCulate<1|2>:PSEarch|PEAKsearch:AUTO
For details refer to "CALCulate<1|2>:ESPectrum:PSEarch|:PEAKsearch:AUTO" on page 6.24.
This command is available from firmware version 1.80.
CALCulate<1|2>:PSEarch|PEAKsearch:MARGin
For details refer to "CALCulate<1|2>:ESPectrum:PSEarch|:PEAKsearch:MARGin" on page
6.25.
This command is available from firmware version 1.80.
CALCulate<1|2>:PSEarch|PEAKsearch:PSHow
For details refer to "CALCulate<1|2>:ESPectrum:PSEarch|:PEAKsearch:PSHow" on page 6.24.
This command is available from firmware version 1.80.

Table of Contents