DIAGnostic Subsystem R&S FSL
1300.2519.12 6.124 E-11
D
IAGnostic<1|2>:SERVice:INPut:PULSed:PRATe
This command sets the comb generator frequency. This command only takes effect, if the internal
reference signal is selected for calibration (DIAGnostic<1|2>:SERVice:INPut[:SELect]).
Parameter
COMB1 65.8333 MHz / 1
COMB64 65.8333 MHz / 64
COMB65 65.8333 MHz / 65
Example
DIAG:SERV:INP:PULS:PRAT COMB64
Sets the comb generator frequency to
64
1
of 65.8333 MHz.
Characteristics
*RST value: COMB1
SCPI: conform
Mode
all
DIAGnostic<1|2>:SERVice:NSOurce (option Additional Interfaces, B5)
This command switches the 28 V supply of the noise source at the rear panel on or off.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
Example
DIAG:SERV:NSO ON
Characteristics
*RST value: OFF
SCPI: device-specific
Mode
all