R&S FSL DISPlay Subsystem
1300.2519.12 6.133 E-11
Example
SWE:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE MAXH
Switches on the calculation of the for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
Characteristics
*RST value: WRITe for TRACe1, STATe OFF for TRACe2/3/4/5/6
SCPI: device–specific
Mode
all
DISPlay[:WINDow<1|2>]:TRACe<1...6>:MODE:HCONtinuous
This command defines, whether traces in Min Hold, Max Hold and Average mode are reset after
parameter change or not.
Normally, the measurement is started anew after parameter changes, before the measurement
results are evaluated (e.g. using a marker). In all cases that require a new measurement after
parameter changes, the trace is reset automatically to avoid false results (e.g. with span
changes). For applications that require no reset after parameter changes, the automatic reset
can be switched off.
The numeric suffixes <1|2> are irrelevant.
Parameter
OFF After certain parameter changes the traces are reset.
ON The automatic reset is switched off.
Example
DISP:WIND:TRAC3:MODE:HCON ON
Switches off the reset function.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A, CATV
DISPlay[:WINDow<1|2>]:TRACe<1...6>:X:SPACing
This command toggles between linear and logarithmic scaling of the x-axis.
Parameter
LINear | LOGarithmic
Example
DISP:TRAC:X:SPAC LOG
Sets the x-axis to a logarithmic scale